OptiLASIK(TM) Laser Vision Correction*

Are you a prospective LASIK patient?

Alcon

Change text size: A | A | A

Technical Data

ALLEGRO Analyzer® Device

Type Diagnostic laser device
Laser class Class 2 (IEC 60825), class II (21 CFR 1040)
Working principle Tscherning principle. A light bundle is projected onto the retina and captured by a light sensitive camera. The distortion of the light bundle in relation to the reference light bundle is calculated and the wavefront error displayed.
Resolution 13 x 13 dots (168 as the central point is not measured)
Measuring beam Diode 660 nm (visible light), laser class 2 (IEC 60825)
Measurable pupil diameter 3 to 8 mm
Measurement range -12 to +6 D sphere / -6 to +6 D astigmatism
Centration Eye tracking function (centered on the gravity center of the pupil)
Accuracy of the refraction +/- 0.01 D measurement
Repeatability +/- 0.05 D in consecutive measurements
Accuracy of the wavefront measurements Zernike polynomials up to the 6th order
Focusing IR illumination of the cornea, reflection is detected by an IR sensitive video camera
Acquisition time app. 250 ms
Operation Data input and data output via PC, patient acquisition with XYZ stage
Capturing Manual or automatic / Automatic capturing if the pupil in x/y-axis is within +/- 100 µm and in the z-axis within +/- 200 µm
Calibration Calibration with calibrated test eye.
Verification of the complete measurement process at any time.
Cooling Ambient air
Temperature range 18 – 30 °C (65 – 86 °F)
Humidity 30 – 80 % at 25 °C (77 °F), not condensing
Mains connection 230 V / 16 A / 50 Hz, 110 V / 16 A / 60 Hz
Power consumption 250 W
Dimensions Optical unit (L x W x H): 45 x 21 x 50 cm (17.7 x 8.3 x 19.7 inches)
Horizontal traverse path 11 cm (4.3 inches)
Longitudinal traverse path 9 cm (3.5 inches)
Weight 15 kg (optical unit), 33 pounds (optical unit)
Electrical safety class Class I, type B (IEC 60601)
Medical device class Class IIb product, Medical Device Directive (MDD) 93/42/EEC
Device is in compliance with IEC 60601, IEC 60825, Medical Device Directive (MDD) 93/42/EEC, 21 CFR 1040.10, 21 CFR 1040.11

Specifications are subject to change.